Electrohms (offers) support  for all products  in manufacture range at present and also provides  customized  products  from the design stage to manufacture .

Energize the sensor at the rated input voltage and observe the output :

Case 1:  closed loop the offset current should be measured . The values should be  in micro amps or milli amps as specified in our data sheet.

Case 2:  open loop the offset output voltage should be measured. The values should be in milli volt as specified in our data sheet.